B-138 A Harmonization Program Leading to Improved Accuracy and Precision of Hemoglobin A1c (HbA1c) Testing: A Six-Year Multi-Site, Multi-Instrument Review and Comparison
Abstract Background Across four global laboratory locations in North America, Europe, and Asia a quality control system was implemented in the measurement of Hemoglobin A1c (HbA1c). The intent was to improve accuracy and reduce imprecision, both being independent contributors to total analytical per...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 2024-10, Vol.70 (Supplement_1) |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Background
Across four global laboratory locations in North America, Europe, and Asia a quality control system was implemented in the measurement of Hemoglobin A1c (HbA1c). The intent was to improve accuracy and reduce imprecision, both being independent contributors to total analytical performance.
Methods
Quality control (QC) data, created during routine clinical operations, was used for this analysis. Analysis of QC data evaluated bias and precision separately and provided the ability to predict and monitor performance. The in-process harmonization program indicated a change in instrumentation was needed to meet and maintain a more stringent analytical error profile. The instrumentation change allowed for a before and after comparison across platforms.
Results
Use of a harmonized quality control system allowed for the visualization of the role bias and precision play in measurements. Improved performance was confirmed using native matrix samples tested routinely from participation in the National Glycohemoglobin Standardization Program. Independently assessing bias and precision highlighted the cumulative and offsetting effects that challenge quality assessment and risk mitigation. Bias appeared to have both a transient and systematic nature. The transient bias contributed to the overall normal distributive property of the quality data. In contrast, the systematic accumulation of bias appeared to occur slowly, over multiple reagent and calibration cycles. The accumulation of bias occurred in the direction of the assay drift. While precision commonly consumes the largest amount of analytical error within an instrument, the dual (negative and positive) nature of bias similarly affects the harmonization of an instrument group.
Conclusions
Changing the way data is viewed to include assessments of bias improved performance and aided in the identification of opportunities to reduce analytical variances that results in inter- and intra-individual variation. Harmonization of instruments should include methods to assess and describe bias, independent of precision. |
---|---|