A-308 Placental Growth Factor Kryptor Immunoassay Reveals Quality Control Imprecision and Drift
Abstract Background The utility and value of determining the soluble fms-like tyrosine kinase-1 to placental growth factor (sFlt-1/PlGF) ratio for predicting the likelihood of progression to pre-eclampsia in pregnant women admitted with hypertension are increasingly becoming widely recognized. A key...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 2024-10, Vol.70 (Supplement_1) |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Background
The utility and value of determining the soluble fms-like tyrosine kinase-1 to placental growth factor (sFlt-1/PlGF) ratio for predicting the likelihood of progression to pre-eclampsia in pregnant women admitted with hypertension are increasingly becoming widely recognized. A key finding of the multi-center PRAECIS trial (NEJM Evid 2022;1(12)) is the observation that sFlt-1/PlGF ratios ≥ 40 correlate with higher risk for adverse maternal outcomes. The sFlt-1 and PlGF immunoassays performed on the ThermoFisher Scientific B·R·A·H·M·S Kryptor autoanalyzer became the first FDA-approved tests in 2023 for determining sFlt-1/PlGF ratio. In this study, an evaluation of the analytical performance of the sFlt-1 and PlGF KRYPTOR immunoassays was performed, which included analyte measuring range (AMR), accuracy, precision, and stability studies.
Methods
Calibration and QC materials (consisting of lyophilized recombinant human proteins), test reagents and solutions, and consumables were purchased and provided by ThermoFisher Scientific. Analytical runs were performed according to vendor’s instructions by trained medical laboratory scientists. Between-day precision of our two instruments for both immunoassays was assessed by assaying all three levels of sFlt-1 and PlGF QC materials, as well as patient serum pools with low PlGF concentration, once to twice daily for multiple consecutive days.
Results
While the analytical performance of the sFlt-1 immunoassay was robust and performed according to expectation (within-day CV |
---|---|