A-080 Morning Cortisol for Rule-in and Rule-out of Adrenal Insufficiency in an Asian Population
Abstract Background Short ACTH stimulation (Synacthen) tests are performed to investigate for adrenal insufficiency. Previous work suggests that morning cortisol concentrations can be used to rule in or out adrenal insufficiency to avoid performance of a Synacthen test. This study aimed to determine...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 2024-10, Vol.70 (Supplement_1) |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Background
Short ACTH stimulation (Synacthen) tests are performed to investigate for adrenal insufficiency. Previous work suggests that morning cortisol concentrations can be used to rule in or out adrenal insufficiency to avoid performance of a Synacthen test.
This study aimed to determine the optimal cortisol concentration for rule-in or rule-out adrenal insufficiency based on 30-minute cortisol (COR30) for adequate response using the assay-specific cut-off from The Endocrine Society of Australia/The Australasian Association for Clinical Biochemistry and Laboratory Medicine/The Royal College of Pathologists of Australasia (ESA/AACB/RCPA) Harmonization of Endocrine Dynamic Testing - Adult (HEDTA), in an Asian population.
Methods
Synacthen tests performed at our institution from 2012-2024, with a baseline cortisol between 6 and 10 am (COR0(6-10)), were extracted from the laboratory database. Plasma cortisol concentrations were measured by Beckman Coulter DxI 800. Synacthen tests were classified as adequate or inadequate based on the Beckman Coulter specific HEDTA cut-off for COR30 of 420 nmol/L. Performance characteristics (sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and area under receiver operator characteristic (AUROC) curve) of COR0(6-10) concentrations to predict adequate or inadequate response were calculated using Microsoft Excel.
Results
4609 Synacthen tests were included, 77% of which had adequate Synacthen response. COR0(6-10) >300 nmol/L could exclude inadequate COR30 response with sensitivity 95%, specificity 48%, PPV 35%, NPV 97%; and COR0(6-10) >364 nmol/L had sensitivity 99%, specificity 29%, PPV 29%, NPV 99%. COR0(6-10) |
---|---|