A-120 Performance Evaluation of Three Sentinel Diagnostics’ Assays on Roche Cobas c 303 Analytical Unit
Abstract Objective The aim of this study is to evaluate the analytical performances of the following Sentinel Diagnostics’ assays: Total Bile Acid Liquid, Copper, Aldolase, on Roche cobas c 303 analytical unit. Methods Performance evaluation included Limit of Blank (LoB), Limit of Detection (LoD), L...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 2023-09, Vol.69 (Supplement_1) |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Objective
The aim of this study is to evaluate the analytical performances of the following Sentinel Diagnostics’ assays: Total Bile Acid Liquid, Copper, Aldolase, on Roche cobas c 303 analytical unit.
Methods
Performance evaluation included Limit of Blank (LoB), Limit of Detection (LoD), Limit of Quantitation (LoQ), linearity, intra assay precision, inter assay precision, on board reagent stability, prozone effect, instrument correlation following the current CLSI guidelines protocols. Data were evaluated using Microsoft Excel statistical tool Analyse-it. Analytical Measuring Range (AMR) was defined through linearity and sensitivity.
Results
Summary of the analytical performances for all the assays tested is reported in table below.
Conclusion
All assays showed good results for all the parameters tested on Roche cobas c 303 analytical unit making these kits very suitable for a routine measurement of these analytes.
Parameters
Total bile acids
Copper
Aldolase
Sensibility—LOB
≤0.2 μmol/L
≤3.0 μg/dL
≤1.0 U/L
Sensibility—LOD
≤0.5 μmol/L
≤5.0 μg/dL
≤2.0 U/L
Sensibility—LOQ
≤1.0 μmol/L
≤7.0 μg/dL
≤3.0 U/L
Linearity
up to 200 μmol/L
up to 500 μg/dL
up to 30 U/L
Precision—Intra assay
%CV ≤ 0.8%
%CV ≤ 1.7%
%CV ≤ 4.1%
Precision—Inter assay
%CV ≤ 3.2%
%CV ≤ 2.6%
%CV ≤ 4.1%
On board reagent stability
up to 60 days
up to 60 days
up to 30 days
No prozone effect up to
820 μmol/L
1500 μg/dL
90 U/L
Correlation vs cobas c503 analytical unit
n = 110 slope = 0.99 intercept = 0.07 μmol/L r = 0.994
n = 110 slope = 1.00 intercept = 1.94 μg/dL r = 1.000
n = 110 slope = 1.04 intercept = 0.29 U/L r = 0.998 |
---|---|