Application of an Experimental Design for the Optimization and Validation of a New HPLC Method for the Determination of Vancomycin in an Extemporaneous Ophthalmic Solution

An experimental design has been used to develop and optimize a new high-performance liquid chromatographic (HPLC) method for the determination of Vancomycin in an extemporaneous ophthalmic solution. After the preliminary studies and literature review, the optimized method was carried out on a second...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of chromatographic science 2008-10, Vol.46 (9), p.828-834
Hauptverfasser: Enrique, Montse, García-Montoya, Encarna, Miñarro, Montserrat, Orriols, Anna, Ticó, Joseph Ramon, Suñé-Negre, Joseph Maria, Pérez-Lozano, Pilar
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An experimental design has been used to develop and optimize a new high-performance liquid chromatographic (HPLC) method for the determination of Vancomycin in an extemporaneous ophthalmic solution. After the preliminary studies and literature review, the optimized method was carried out on a second generation of a C18 reverse-phase column (Luna 150 × 4.6 mm i.d., 5 µm particle size) and using methanol as organic phase, a less toxic solvent than acetonitrile, described in the extended literature. The experimental design consisted of a Placket-Burman design where six different variables were studied (flow rate, mL/min; temperature, °C; pH mobile phase; % buffer solution; wavelength; and injection volume) to obtain the best suitability parameters (Capacity factor-K', tailing factor, resolution, and theoretical plates). After the optimization of the chromatographic conditions and statistical treatment of the obtained results, the final method uses a mixture of a buffer solution of water—phosphoric acid (85%) (99.83:0.17, v/v) adjusted to pH 3.0 using triethylamine and mixed with methanol (87:13, v/v). The separation is achieved using a flow rate of 1.0 mL/min at 35°C. The UV detector was operated at 280 nm. The validation study carried out, demonstrates the viability of the method, obtaining a good selectivity, linearity, precision, accuracy, and sensitivity.
ISSN:0021-9665
1945-239X
DOI:10.1093/chromsci/46.9.828