Evaluation of Roche Hemolysis Index for Reporting of Plasma Hemoglobin, and Considerations Regarding Hemolysis Detection in Use of Ventricular-Assist Devices
Abstract Introduction/Objective Our cardiac intensive care unit (CICU) asked the laboratory to provide plasma hemoglobin testing to monitor effects of ventricular-assist devices (VAD). Roche hemolysis index (HI) has been reported to be suitable for this purpose. Use of HI as a reportable test, howev...
Gespeichert in:
Veröffentlicht in: | American journal of clinical pathology 2021-10, Vol.156 (Supplement_1), p.S31-S31 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Introduction/Objective
Our cardiac intensive care unit (CICU) asked the laboratory to provide plasma hemoglobin testing to monitor effects of ventricular-assist devices (VAD). Roche hemolysis index (HI) has been reported to be suitable for this purpose. Use of HI as a reportable test, however, must be validated in-house as a laboratory- developed test (LDT), according to CLIA regulations. We describe results of our evaluation, and caveats for intended use.
Methods/Case Report
Concentrated hemolysate was produced by osmotic lysis of packed red blood cells in water. Hemolysis in plasma was by dilution of concentrated hemolysate. Verification of HI as a measure of hemoglobin (mg/dL) used the Sigma-Aldrich spectrophotometric hemoglobin assay as a gold standard. Linearity, linear range, sensitivity, were investigated by dilution measurements; interferences by admixture experiments; reproducibility (precision, intra- and inter-assay) in each case by replicate measurements (n = 20).
Results (if a Case Study enter NA)
HI was confirmed to correspond to hemoglobin in mg/dL. Linearity was between 2-1000 mg/dL (r2 >0.99). Intra-assay precisions were |
---|---|