Reading of gafchromic EBT-3 film using an overhead scanner

Gafchromic film, a commercially available radiochromic film, has been developed and widely used as an effective tool for radiation dose verification and quality assurance in radiotherapy. However, the orientation effect in scanning a film remains a concern for practical application in beam profile m...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomedical physics & engineering express 2024-07, Vol.10 (5), p.55004
Hauptverfasser: Bantan, H, Yasuda, H
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Gafchromic film, a commercially available radiochromic film, has been developed and widely used as an effective tool for radiation dose verification and quality assurance in radiotherapy. However, the orientation effect in scanning a film remains a concern for practical application in beam profile monitoring. To resolve this issue, the authors introduced a novel method using an overhead scanner (OHS) coupled with a tracing light board instead of a conventional flatbed scanner (FBS) to read Gafchromic EBT3 films. We investigated the orientation effect of the EBT3 film with a regular hexagonal shape after irradiation with 5 Gy x-rays (160 kV, 6.3 mA) and compared the digitized images acquired using a commercially available OHS (CZUR Aura) and a conventional FBS (EPSON GT-X980). As a result, RGB color intensities acquired from the OHS showed significantly lower orientation effect of the color intensities of RGB components than those from FBS. This finding indicates the high potential of the proposed method for achieving more precise two-dimensional dosimetry. Further studies are required to confirm the effectiveness of this method under different irradiation conditions over a wider dose range.
ISSN:2057-1976
2057-1976
DOI:10.1088/2057-1976/ad5cf8