Development of a dosimeter prototype with machine learning based 3-D dose reconstruction capabilities

A 3-D dosimeter fills the need for treatment plan and delivery verification required by every modern radiation-therapy method used today. This report summarizes a proof-of-concept study to develop a water-equivalent solid 3-D dosimeter that is based on novel radiation-hard scintillating material. Th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomedical physics & engineering express 2022-01, Vol.8 (1), p.15009
Hauptverfasser: Finneman, G M, Eichhorn, O H, Meskell, N R, Caplice, T W, Benson, A D, Abu-Halawa, A S, Ademoski, G L, Clark, A C, Gayer, D S, Hendrickson, K N, Debbins, P A, Onel, Y, Ayan, A S, Akgun, U
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A 3-D dosimeter fills the need for treatment plan and delivery verification required by every modern radiation-therapy method used today. This report summarizes a proof-of-concept study to develop a water-equivalent solid 3-D dosimeter that is based on novel radiation-hard scintillating material. The active material of the prototype dosimeter is a blend of radiation-hard peroxide-cured polysiloxane plastic doped with scintillating agent P-Terphenyl and wavelength-shifter BisMSB. The prototype detector was tested with 6 MV and 10 MV x-ray beams at Ohio State University's Comprehensive Cancer Center. A 3-D dose distribution was successfully reconstructed by a neural network specifically trained for this prototype. This report summarizes the material production procedure, the material's water equivalency investigation, the design of the prototype dosimeter and its beam tests, as well as the details of the utilized machine learning approach and the reconstructed 3-D dose distributions.
ISSN:2057-1976
2057-1976
DOI:10.1088/2057-1976/ac396c