Correction method for 3D non-linear drift distortions in atomic force microscopy raster measurements

A method to correct non-linear drift distortions in all three coordinate axes of atomic force microscope (AFM) images is presented. The method uses two measurements of the sample with two fast scan axes orthogonal to each other. Both AFM images are divided into segments and the shifts of the surface...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Measurement science & technology 2023-02, Vol.34 (2), p.25001
Hauptverfasser: Degenhardt, Johannes, Tutsch, Rainer, Dai, Gaoliang
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method to correct non-linear drift distortions in all three coordinate axes of atomic force microscope (AFM) images is presented. The method uses two measurements of the sample with two fast scan axes orthogonal to each other. Both AFM images are divided into segments and the shifts of the surface features of the segments of both images are determined. From these shifts subsequently the drift of both measurements is calculated. Depending on the segments used, significant non-linearities of the drift can be corrected. The two required measurements for this method do not have to be carried out in direct succession. With this method it is therefore possible to correct drift in an existing AFM image by measuring the sample again later. Although the method has been developed for AFM, it can also be used for other scanning probe microscopes.
ISSN:0957-0233
1361-6501
DOI:10.1088/1361-6501/ac9992