Influence of the simultaneous calibration of multiple ring dosimeters on the individual absorbed dose
Ring dosemeters for personal dosimetry are calibrated in accredited laboratories following ISO 4037-3 guidelines. Multiple irradiation of dosemeters would save time, but has to be carefully studied, since the scattering conditions could change and influence the absorbed dose in nearby dosimeters. Mo...
Gespeichert in:
Veröffentlicht in: | Journal of radiological protection 2021-06, Vol.41 (2), p.384-397 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Ring dosemeters for personal dosimetry are calibrated in accredited laboratories following ISO 4037-3 guidelines. Multiple irradiation of dosemeters would save time, but has to be carefully studied, since the scattering conditions could change and influence the absorbed dose in nearby dosimeters. Monte Carlo simulations using PENELOPE-2014 were performed to explore the need of increasing the uncertainty of Hp(0.07) in the simultaneous irradiation of three and five DXT-RAD 707H-2 (Thermo Scientific) ring dosemeters with beam qualities: N-30, N-80 and N-300. Results show that the absorbed dose in each dosemeter is compatible with each other and with the reference simulation (a single dosemeter) with a coverage probability of 95% (k = 2). Comparison with experimental data yielded consistent results with the same coverage probability. Therefore, five ring dosemeters can be simultaneously irradiated with beam qualities ranging, at least, between, N-30 and N-300 with a negligible impact on the uncertainty of Hp(0.07). |
---|---|
ISSN: | 0952-4746 1361-6498 |
DOI: | 10.1088/1361-6498/abef53 |