Effects of applied contact force and volume control setting on output force levels of the BAHA® Softband: Efectos de la fuerza aplicada de contacto y de los ajustes en el control de volumen sobre los niveles de fuerza de salida del BAHA® Softband

The BAHA® Softband has been developed to provide a transcutaneous anchor for a BAHA® until a child is a surgical candidate for the percutaneous BAHA® implant. We tested the objective output force level of the BAHA® Classic 300 and Compact connected to a Softband on an artificial mastoid to determine...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of audiology 2006-01, Vol.45 (5), p.301-308
Hauptverfasser: Hodgetts, William E., Scollie, Susan D., Swain, Ryan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The BAHA® Softband has been developed to provide a transcutaneous anchor for a BAHA® until a child is a surgical candidate for the percutaneous BAHA® implant. We tested the objective output force level of the BAHA® Classic 300 and Compact connected to a Softband on an artificial mastoid to determine: (1) the effects of direct contact force on output force levels (dB); and (2) the required volume control setting to ensure audibility of speech (assuming an average adult reference equivalent threshold force level). Direct contact force was varied from 2 to 5 N in 1 N steps. Output force level increased with increasing contact force. However, the average increase was 3 dB or lower, suggesting that the contact force is of minor importance. Volume control setting appears to be of much greater importance. It is suggested that the volume setting of either device be set to at least 2.5 to ensure audibility of conversational speech. Data from normal-hearing adults with simulated conductive hearing losses are presented to validate this conclusion.
ISSN:1499-2027
1708-8186
DOI:10.1080/14992020600582133