Operational Characteristics of the Evaporative Light Scattering Detector Used in Analysis of Ethoxylated Alcohols
Influence of the temperature of a drift tube and flow rate of nitrogen as nebulizing gas with 100% hexane and 40% hexane plus 60% (5% H 2 O in 2-propanol) as mobile phases on the respones of ELS detector for selected standards of ethoxylated alcohols was examined. For optimal conditions (1.5 dm 3 N...
Gespeichert in:
Veröffentlicht in: | Journal of liquid chromatography & related technologies 1996-04, Vol.19 (7), p.1013-1032 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Influence of the temperature of a drift tube and flow rate of nitrogen as nebulizing gas with 100% hexane and 40% hexane plus 60% (5% H
2
O in 2-propanol) as mobile phases on the respones of ELS detector for selected standards of ethoxylated alcohols was examined. For optimal conditions (1.5 dm
3
N
2
/min, 50°C) calibration curves for both mobile phases were created. The response of ELSD for standards not undergoing evaporation is mainly affected by the composition of the mobile phase. |
---|---|