Radial Differential Mobility Analyzer for One Nanometer Particle Classification

We present details on the design, construction, and calibration of a radial differential mobility analyzer suitable for sizing aerosol nanoparticles with mobility diameters as small as one nanometer. Design improvements targeted decreasing the residence time with a concomitant reduction in particle...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Aerosol science and technology 2009-01, Vol.43 (1), p.53-59
Hauptverfasser: Brunelli, N. A., Flagan, R. C., Giapis, K. P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present details on the design, construction, and calibration of a radial differential mobility analyzer suitable for sizing aerosol nanoparticles with mobility diameters as small as one nanometer. Design improvements targeted decreasing the residence time with a concomitant reduction in particle agglomeration and diffusional losses to walls. Calibration was performed using aerosolized monodisperse molecular ions. Instrument resolution was found to increase with applied voltage as V 1/2 , indicating that diffusion is broadening the transfer function and reaching a value of ∼7 for the largest molecular ion measured (mobility diameter of 1.47 nm). The mobilities of three molecular ions in the tetra-alkyl ammonium halide homologous series are also reported.
ISSN:0278-6826
1521-7388
DOI:10.1080/02786820802464302