Liquid Chromatographic Determination of Cyclosporine-A in Blood, with Chromosorb P Columns Used for Sample Purification
A fast liquid-chromatographic determination of whole-blood cyclosporine A concenration is described. A sample preparation consisting of diethyl ether extraction and Chromosorb P column purification of extract, requires 1 mL of whole blood and takes 10 min of technical effort per sample. A reversed-p...
Gespeichert in:
Veröffentlicht in: | Journal of liquid chromatography 1991-06, Vol.14 (9), p.1683-1690 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A fast liquid-chromatographic determination of whole-blood cyclosporine A concenration is described. A sample preparation consisting of diethyl ether extraction and Chromosorb P column purification of extract, requires 1 mL of whole blood and takes 10 min of technical effort per sample. A reversed-phase C
1
column (5-μm particles) is used, with acetonitrile/methanol/water (20/45/35 by vol) as mobile phase. Cyclosporine A is quantified by absorbance at 214 nm, with cyclosporine D as internal standard. Chromatographic development is complete in 8 min. Linearity is verified by a five-point calibration curve in the range 50-900 μg/L, correlation coefficient r>0.998, y = 0.001x − 0.053. Lower limit of sensitivity is 25 μg/L. Extraction efficiency was over 70%, accuracy varied between −7.1% and +3.3%s, CVs were |
---|---|
ISSN: | 0148-3919 2331-0413 |