Design and validation of microfluidic parameters of a microfluidic chip using fluid dynamics

The internal fluidic parameters of microfluidic channels must be analyzed to solve fundamental microfluidic problems, including microscale transport problems involving thermal analysis, chemical reactivity, velocity, pressure drop, etc., for developing good-quality chemical and biological products....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:AIP advances 2021-07, Vol.11 (7), p.075224-075224-9
Hauptverfasser: Ahmed, Feroz, Yoshida, Yuichi, Wang, Jin, Sakai, Kenji, Kiwa, Toshihiko
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The internal fluidic parameters of microfluidic channels must be analyzed to solve fundamental microfluidic problems, including microscale transport problems involving thermal analysis, chemical reactivity, velocity, pressure drop, etc., for developing good-quality chemical and biological products. Therefore, the characterization and optimization of the interaction of chemical and biological solutions through microfluidic channels are vital for fluid flow design and engineering for quality assurance in microfluidic platforms. As the internal structures and kinetics of microfluidic channels are becoming increasingly complex, experiments involving optimal fluidic and transport designs are challenging to perform with high accuracy. However, highly integrated simulation tools can guide researchers without specialized computational fluid backgrounds to design numerical prototypes of highly integrated devices. In this study, a microfluidic chip with two inlet wells and one outlet well was fabricated from polydimethylsiloxane following which simulations were performed using an ANSYS Fluent tool influenced by computational fluid dynamics at a nearly identical scale. The pressure drop and velocity profiles of the interaction of two pH buffer solutions (pH 4 and 10) through the designed microfluidic chip were qualitatively estimated from experimental data analysis and validated with the simulation results obtained from the CFD-influenced ANSYS Fluent tool.
ISSN:2158-3226
2158-3226