Planar, strong magnetic field source for a chip ion trap

We present a planar, scalable magnetic field source, originally conceived for a chip ion trap. It consists of two symmetric sections, each with several independent currents arranged in coplanar, concentric rectangular loops. The currents allow for tuning the strength of the field and its lowest-orde...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2020-10, Vol.91 (10), p.103201-103201
Hauptverfasser: Pinder, Jonathan, Lacy, John H., Willetts, Ryan, Cridland Mathad, April, Uribe, Alberto, Verdú, José
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present a planar, scalable magnetic field source, originally conceived for a chip ion trap. It consists of two symmetric sections, each with several independent currents arranged in coplanar, concentric rectangular loops. The currents allow for tuning the strength of the field and its lowest-order derivatives at one discretional position along the source’s vertical symmetry axis, a few mm above its surface. We describe the construction and calibration of the device and the cryogenic setup. The two most important current configurations for a Penning ion trap, the homogeneous field and the magnetic bottle, are investigated experimentally. Homogeneous fields around 0.5 T are routinely reached. We discuss the maximum attainable field, and we briefly describe ongoing further developments aiming at homogeneous fields well above 1 T.