A new fast response cryogenic evaporative calorimeter
We present the principle and implementation of a new type of fast response evaporative calorimeter designed to work at cryogenic temperatures and above-ambient pressures. It is capable of measuring input energy from an electric pulse and the thermal output energy by measuring the evaporation of liqu...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2020-08, Vol.91 (8), p.085103-085103 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present the principle and implementation of a new type of fast response evaporative
calorimeter designed to work at cryogenic temperatures and above-ambient pressures. It is
capable of measuring input energy from an electric pulse and the thermal output energy by
measuring the evaporation of liquid nitrogen through a mass flow meter. This system may be
used to measure either the steady heat output from the system submersed under the cryogen
or the heat output that results from a fast square-wave profile electrical pulse of
duration from 10 µs or longer. The energy output of metal capillary-wire
composite systems has been measured calorimetrically. A four-wire measurement was used to
monitor the input electric energy with an uncertainty less than 5% for a typical pulse.
Mass flow meters and pressure regulation systems were used to monitor the rate of
evaporation of liquid nitrogen with a typical precision of 2 std.-ml/min. For a typical
pulse, the integrated mass flow of nitrogen could be determined with an uncertainty less
than 3%. The pressure controllers and ballast compliance volumes allow the system to
return to a steady state of mass flow in less than 2 min following an electric pulse. The
system is capable of housing and measuring four separate wire-capillary systems in a
single Dewar. On average, a calibration resulted in 3.9 std. ml evaporated per joule of
input energy. This corresponds to a 97% efficiency for this calorimeter. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/5.0013713 |