An ultra-low detection-limit optofluidic biosensor with integrated dual-channel Fabry-Pérot cavity
A silicon-on-insulator based optofluidic biosensor with integrated dual-channel Fabry–Pérot cavity is proposed for optical differential detection. A detection limit of 5.5 × 10−8 refractive index unit is experimentally demonstrated, owing to the high quality factor of the cavity and the differential...
Gespeichert in:
Veröffentlicht in: | Applied physics letters 2013-04, Vol.102 (16) |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A silicon-on-insulator based optofluidic biosensor with integrated dual-channel Fabry–Pérot cavity is proposed for optical differential detection. A detection limit of 5.5 × 10−8 refractive index unit is experimentally demonstrated, owing to the high quality factor of the cavity and the differential detection, which can extract the small signal for efficient amplification and greatly reduce the system noise. Moreover, the measurement system features low cost compared with that of surface-plasmon-resonance sensor and ring-resonator sensor. |
---|---|
ISSN: | 0003-6951 1077-3118 |
DOI: | 10.1063/1.4802805 |