Internal losses in polyvinylidene fluoride (PVF2) ultrasonic transducers

The dielectric and acoustic losses in oriented PVF2 have been measured from 0.1 MHz to 100 MHz and from 1.5 MHz to 39 MHz, respectively, in the direction perpendicular to the stretching direction at room temperature. A broad absorption peak located at around 5 MHz is observed in each case. The locat...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of applied physics 1979-01, Vol.50 (12), p.8031-8033
Hauptverfasser: Leung, W. P., Yung, K. K.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The dielectric and acoustic losses in oriented PVF2 have been measured from 0.1 MHz to 100 MHz and from 1.5 MHz to 39 MHz, respectively, in the direction perpendicular to the stretching direction at room temperature. A broad absorption peak located at around 5 MHz is observed in each case. The locations of these absorption peaks are of great disadvantage for ultrasonic transducer fabrication. The longitudinal acoustic velocity in the direction perpendicular to the stretching direction is also measured from 1.5 MHz to 39 MHz at room temperature. It is found that the velocity increases monotonically in this frequency range. As a result, the resonance frequency of a PVF2 transducer will be affected proportionally.
ISSN:0021-8979
1089-7550
DOI:10.1063/1.325938