Thermal diagnostics front-end electronics for LISA Pathfinder
Precision temperature measurements are required in the LTP, the LISA technology package, for various diagnostics objectives. In this article, we describe in detail the front-end electronics design and the associated temperature sensors to achieve the LTP requirements: noise equivalent temperature of...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2007-10, Vol.78 (10), p.104904-104904-10 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Precision temperature measurements are required in the LTP, the LISA technology package, for various diagnostics objectives. In this article, we describe in detail the front-end electronics design and the associated temperature sensors to achieve the LTP requirements: noise equivalent temperature of
10
μ
K
Hz
−
1
∕
2
in the frequency range from
1
to
30
mHz
at room temperature. We designed an ac Wheatstone bridge and a subsequent digital demodulation to minimize
1
∕
f
noise. We show experimental results where the required sensitivity in the measurement bandwidth is fulfilled. |
---|---|