A Fabry-Perot interferometer for micrometer-sized cantilevers

We have developed a Fabry-Perot interferometer detecting the deflection of micrometer-sized cantilevers and other micromechanical devices, at a working distance of 0.8 mm . At 1 MHz , a noise floor of 1 fm ∕ Hz is obtained. The detector is mounted on a piezo motor for three-axis alignment. The angul...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied physics letters 2005-02, Vol.86 (7), p.074101-074101-3
Hauptverfasser: Hoogenboom, B. W., Frederix, P. L. T. M., Yang, J. L., Martin, S., Pellmont, Y., Steinacher, M., Zäch, S., Langenbach, E., Heimbeck, H.-J., Engel, A., Hug, H. J.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We have developed a Fabry-Perot interferometer detecting the deflection of micrometer-sized cantilevers and other micromechanical devices, at a working distance of 0.8 mm . At 1 MHz , a noise floor of 1 fm ∕ Hz is obtained. The detector is mounted on a piezo motor for three-axis alignment. The angular alignment is not critical. The interferometer can be operated in vacuum, air, and liquid. It is particularly suited for scanning force microscopy with small cantilevers, or with larger cantilevers simultaneously monitoring vertical and lateral forces.
ISSN:0003-6951
1077-3118
DOI:10.1063/1.1866229