KB–PJX—A streaked imager based on a versatile x-ray microscope coupled to a high-current streak tube (invited)

A re-entrant, highly adaptable, x-ray streaked imager has been developed for OMEGA to increase the sensitivity and spatial resolution in hydrodynamic-stability experiments. It is based on a four-mirror Kirkpatrick–Baez (KB) microscope, coupled to a high-current streak tube. The unique mechanical ass...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of Scientific Instruments 2004-10, Vol.75 (10), p.4063-4068
Hauptverfasser: Gotchev, O. V., Jaanimagi, P. A., Knauer, J. P., Marshall, F. J., Meyerhofer, D. D.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A re-entrant, highly adaptable, x-ray streaked imager has been developed for OMEGA to increase the sensitivity and spatial resolution in hydrodynamic-stability experiments. It is based on a four-mirror Kirkpatrick–Baez (KB) microscope, coupled to a high-current streak tube. The unique mechanical assembly of the KB optic allows a choice between single- or multi-image modes and one- or two-dimensional imaging. Currently, the optic uses an Ir coating at a grazing angle of 2.1° . The incidence angle has been optimized to maximize throughput in the chosen energy band, centered at 1.5  keV with a full width at half maximum of about 0.4  keV . A calculated resolution of better than 5 μ m over the central 200 μ m of the field of view was verified in inertial confinement fusion experiments. New multilayer mirror elements for high-energy or multiband imaging, take advantage of the flexible mechanical design. Some important features of the PJX streak tube are described.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1786649