Compact high-resolution ion Doppler spectrometer for quartz ultraviolet line emissions

A high-resolution quartz ultraviolet spectrometer based on a single compact 16-channel multianode photomultiplier tube (MAPMT) has been developed and tested on coaxial helicity injection and ohmic transformer current drive plasmas generated on the Helicity Injected Torus at the University of Washing...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2004-05, Vol.75 (5), p.1337-1340
Hauptverfasser: Gu, Peimin, Nagata, Masayoshi, McCollam, Karsten J., Jarboe, Thomas R., Nelson, Brian A., Redd, Aaron J.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A high-resolution quartz ultraviolet spectrometer based on a single compact 16-channel multianode photomultiplier tube (MAPMT) has been developed and tested on coaxial helicity injection and ohmic transformer current drive plasmas generated on the Helicity Injected Torus at the University of Washington. The use of the compact MAPMT allows a simple design with high light collection and high resolution for an ion Doppler shift diagnostic. The high quality and low cost of the design make it uniquely attractive for university scale plasma experiments. The Doppler shift of impurity ion emission lines from two chords of equal impact parameter but of opposite rotational sense is used to determine chord-averaged toroidal flow velocity. The width of the Doppler broadening is used to calculate the ion temperature. A detailed description of the ion Doppler spectrometer measurement system and some typical results are presented.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1711157