Piezoelectric copolymer hydrophones for ultrasonic field characterization
Hydrophones to be used in the characterization of medical ultrasonic transducers have been fabricated using a new polyvinylidene fluoride/trifluoroethylene (VF2/VF3) copolymer. The copolymer has an advantage over VF2 in that it does not require prestretching before poling. Thin copolymer films can b...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 1991-01, Vol.62 (1), p.203-207 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Hydrophones to be used in the characterization of medical ultrasonic transducers have been fabricated using a new polyvinylidene fluoride/trifluoroethylene (VF2/VF3) copolymer. The copolymer has an advantage over VF2 in that it does not require prestretching before poling. Thin copolymer films can be cast from solution and then poled using the corona discharge method. As there is a need for small‐diameter hydrophones to provide good spatial resolution in measuring highly focused ultrasonic beams, hydrophones with diameter as small as 0.1 mm have been made. Both needle‐type and line hydrophones have been tested and their performance reported. In the case of line hydrophones, the output signal is proportional to the line integral of the acoustic pressure and a computer tomographic technique has been used to reconstruct the beam profiles. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.1142306 |