Observations on the use of a thermal conductivity cell to measure the para hydrogen concentration in a mixture of para and ortho hydrogen gas

This paper describes the use of a thermal conductivity cell to measure the para hydrogen concentration in a mixture of para and ortho hydrogen gas. The device was calibrated over its entire range and the output compared with the theoretically expected form. The output of this device had been previou...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 1987-01, Vol.58 (1), p.83-85
Hauptverfasser: Bradshaw, T. W., Norris, J. O. W.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper describes the use of a thermal conductivity cell to measure the para hydrogen concentration in a mixture of para and ortho hydrogen gas. The device was calibrated over its entire range and the output compared with the theoretically expected form. The output of this device had been previously assumed to be linear, this assumption can lead to systematic errors of up to 5.6% in the determination of the para hydrogen concentration. A simple calibration procedure is described.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1139519