A multichannel ion mass spectrometer in laser‐produced plasma studies

The theory and operation of a time‐of‐flight mass spectrometer employing a multichannel electrostatic analyzer are discussed with emphasis placed on laser‐produced plasma studies. It is shown that the finite length of the analyzer leads to enhanced mass resolution. A method of i n s i t u calibratio...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Rev. Sci. Instrum.; (United States) 1976-05, Vol.47 (5), p.548-552
1. Verfasser: Goforth, R. R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The theory and operation of a time‐of‐flight mass spectrometer employing a multichannel electrostatic analyzer are discussed with emphasis placed on laser‐produced plasma studies. It is shown that the finite length of the analyzer leads to enhanced mass resolution. A method of i n s i t u calibration of the analyzer channels is described, and measurements of the resolution and beam divergence are presented. Use of the mass spectrometer to calibrate ion‐collecting probes for mass and kinetic energy measurements is also considered.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1134688