The Press-Fit Condylar total knee system

A total of 108 consecutive Press-Fit Condylar total knee arthroplasties were performed in 94 patients. All patients had implantation with a cemented posterior cruciate-retaining design, which included resurfacing of the patella. Mean age at surgery was 70 years (range, 35-87 years). Patients were fo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2000-09, Vol.15 (6), p.698-701
Hauptverfasser: Buehler, Knute O., Venn-Watson, Eric, D'Lima, Darryl D., Colwell, Clifford W.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A total of 108 consecutive Press-Fit Condylar total knee arthroplasties were performed in 94 patients. All patients had implantation with a cemented posterior cruciate-retaining design, which included resurfacing of the patella. Mean age at surgery was 70 years (range, 35-87 years). Patients were followed for a mean of 9 years (range, 8-10 years) with follow-up for all surviving patients. The average postoperative functional knee score was 96 points. Nonprogressive radiolucent lines were present in 59%. One patellar component was radiographically loose. Five knees underwent revision procedures, none for aseptic loosening. Survivorship was 93.4% at 9 years with revision for any reason as the endpoint and 98.7% with aseptic loosening as the endpoint. This knee arthroplasty shows excellent results at 8 to 10 years with no patients lost to follow-up.
ISSN:0883-5403
1532-8406
DOI:10.1054/arth.2000.8189