Evaluation of the use of cold plasma conditions for Fe isotopic analysis via multi-collector ICP-mass spectrometry: effect on spectral interferences and instrumental mass discrimination
The advantages and disadvantages of using cold plasma conditions in combination with both the standard and the high-transmission ('jet') plasma interface and under dry and wet plasma conditions were evaluated in the context of high-precision isotopic analysis of Fe via multi-collector indu...
Gespeichert in:
Veröffentlicht in: | Journal of analytical atomic spectrometry 2017-03, Vol.32 (3), p.538-547 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The advantages and disadvantages of using cold plasma conditions in combination with both the standard and the high-transmission ('jet') plasma interface and under dry and wet plasma conditions were evaluated in the context of high-precision isotopic analysis of Fe
via
multi-collector inductively coupled plasma-mass spectrometry (MC-ICP-MS). When using the standard interface and wet plasma, cold plasma conditions suppressed the occurrence of the polyatomic ions
40
Ar
14
N
+
and
40
Ar
16
O
+
, but not that of
40
Ar
16
O
1
H
+
. Dry plasma conditions efficiently removed the (hydr)oxide ions, but at the cost of a dramatic increase in the prevalence of
40
Ar
14
N
+
. Thus, interference-free
56
Fe/
57
Fe measurement can be performed at low mass resolution using cold plasma conditions and the standard plasma interface. No alleviation of the spectral interferences due to Ar-based ions was observed with the 'jet' interface with cold plasma conditions. With cold plasma conditions, the instrumental mass discrimination affecting the Fe isotope ratios was mitigated compared to that obtained with hot plasma when the standard interface was used (1.4% per amu
versus
3.2% per amu, respectively). This mitigation in instrumental mass discrimination with cold plasma conditions was not observed when the high-transmission interface was used.
The application of cold plasma conditions for isotopic analysis of Fe
via
MC-ICP-MS was evaluated in combination with the standard and the 'jet' interface and with dry and wet plasma conditions. |
---|---|
ISSN: | 0267-9477 1364-5544 |
DOI: | 10.1039/c6ja00428h |