Continuous flow real-time PCR device using multi-channel fluorescence excitation and detection
High throughput automation is greatly enhanced using techniques that employ conveyor belt strategies with un-interrupted streams of flow. We have developed a 'conveyor belt' analog for high throughput real-time quantitative Polymerase Chain Reaction (qPCR) using droplet emulsion technology...
Gespeichert in:
Veröffentlicht in: | Lab on a chip 2014-02, Vol.14 (3), p.562-568 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | High throughput automation is greatly enhanced using techniques that employ conveyor belt strategies with un-interrupted streams of flow. We have developed a 'conveyor belt' analog for high throughput real-time quantitative Polymerase Chain Reaction (qPCR) using droplet emulsion technology. We developed a low power, portable device that employs LED and fiber optic fluorescence excitation in conjunction with a continuous flow thermal cycler to achieve multi-channel fluorescence detection for real-time fluorescence measurements. Continuously streaming fluid plugs or droplets pass through tubing wrapped around a two-temperature zone thermal block with each wrap of tubing fluorescently coupled to a 64-channel multi-anode PMT. This work demonstrates real-time qPCR of 0.1-10 μL droplets or fluid plugs over a range of 7 orders of magnitude concentration from 1 × 10
1
to 1 × 10
7
. The real-time qPCR analysis allows dynamic range quantification as high as 1 × 10
7
copies per 10 μL reaction, with PCR efficiencies within the range of 90-110% based on serial dilution assays and a limit of detection of 10 copies per rxn. The combined functionality of continuous flow, low power thermal cycling, high throughput sample processing, and real-time qPCR improves the rates at which biological or environmental samples can be continuously sampled and analyzed.
We have developed a 'conveyor belt' analog for real-time quantitative polymerase chain reaction (qPCR) in a continuous flow thermocycler. The device integrates droplet emulsion technology with LED and fiber optic fluorescence excitation in conjunction with a continuous flow thermal cycler to achieve real-time fluorescence detection. |
---|---|
ISSN: | 1473-0197 1473-0189 |
DOI: | 10.1039/c3lc51236c |