Enhanced Visual Wireless Electrochemiluminescence Immunosensing of Prostate-Specific Antigen Based on the Luminol Loaded into MIL-53(Fe)-NH 2 Accelerator and Hydrogen Evolution Reaction Mediation
A sensitive prostate-specific antigen (PSA) detection method using a visual-readout closed bipolar electrode (BPE) system has been introduced by integration of hydrogen evolution reaction (HER) in cathodic pole and electrochemiluminescence (ECL) of luminol loaded within the MIL-53(Fe)-NH (L@MIL-53(F...
Gespeichert in:
Veröffentlicht in: | Analytical chemistry (Washington) 2019-05, Vol.91 (9), p.6383-6390 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A sensitive prostate-specific antigen (PSA) detection method using a visual-readout closed bipolar electrode (BPE) system has been introduced by integration of hydrogen evolution reaction (HER) in cathodic pole and electrochemiluminescence (ECL) of luminol loaded within the MIL-53(Fe)-NH
(L@MIL-53(Fe)-NH
) in the anodic pole. The cathode of the BPE was electrochemically synthesized by 3D porous copper foam, followed by decorating with nitrogen-doped graphene nanosheet and ruthenium nanoparticles. As an alternative, we employed carboxylate-modified magnetic nanoparticles (MNPs) for immobilization of the primary antibody (Ab1) and utilized the L@MIL-53(Fe)-NH
conjugated to secondary antibody (Ab2) as a signaling probe and coreaction accelerator. After sandwiching the target PSA between Ab1 and Ab2, the MNP/Ab1-PSA-Ab2/L@MIL-53(Fe) were introduced to a gold anodic BPE. Finally, the resulting ECL of luminol and H
O
at the anodic poles was monitored using a photomultiplier tube (PMT) or digital camera. The PMT and visual (camera)-based detections showed linear responses from 1 pg mL
to 300 ng mL
(limit of detection 0.2 pg mL
) and 5 pg mL
to 200 ng mL
(limit of detection 0.1 pg mL
), respectively. This strategy provides an effective method for high-performance bioanalysis and opens a new door toward the development of the highly sensitive and user-friendly device. |
---|---|
ISSN: | 0003-2700 1520-6882 |
DOI: | 10.1021/acs.analchem.9b01506 |