The evaluation and optimization of calibration methods for low-cost particulate matter sensors: Inter-comparison between fixed and mobile methods

With the development of the air pollution control, the low-cost sensors are widely used in air quality monitoring, while the data quality of these sensors is always the most concern for users. In this study, data from nine air monitoring stations with standard PM instruments were used as reference a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Science of the total environment 2020-05, Vol.715, p.136791-136791, Article 136791
Hauptverfasser: Qin, Xiaoliang, Hou, Lujian, Gao, Jian, Si, Shuchun
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:With the development of the air pollution control, the low-cost sensors are widely used in air quality monitoring, while the data quality of these sensors is always the most concern for users. In this study, data from nine air monitoring stations with standard PM instruments were used as reference and compared with the data of mobile and fixed PM sensors in Jinan, the capital city of Shandong Province, China. Data quality of PM sensors was checked by the cross-comparison among standard method, fixed and mobile sensors. And the impacts of relative humidity and size distribution (PM2.5/PM10) on the performance of PM sensors were evaluated as well. To optimize the calibration method for both fixed and mobile PM sensors, a two-step model was designed, in which the RH and PM2.5/PM10 ratio were both used as input parameters. We firstly calibrated the sensors with five independent models, and then all the calibrated data were linearly fitted by the LR-final model. In comparison with standard instruments, the LR-final model increased the R2 values of the PM2.5 and PM10 measured by fixed sensors from 0.89 and 0.79 to 0.98 and 0.97, respectively. The R2 values of PM2.5 and PM10 measured by the mobile sensors both increased to 0.99 from 0.79 and 0.62. Overall, the two-step calibration model appeared to be a promising approach to solve the poor performance of low-cost sensors. [Display omitted] •Comparison of standard instruments, fixed sensors, and mobile sensors.•The effects of RH and particle size distribution on the PM sensor were studied.•A two-step calibration model was established, including linear and nonlinear models.•The two-step model can effectively solve the poor performance of low-cost sensors.
ISSN:0048-9697
1879-1026
DOI:10.1016/j.scitotenv.2020.136791