Self-optimizing method and software for calibration and mapping of a laser system for laser machining

•A new calibration method and software is proposed for the self-optimizing calibration of a laser system, for both mechanical and optical axes.•The method and software can be used for automated monitoring and recalibration of the optical axis of a laser system to ensure and maintain a high accuracy....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Optics and lasers in engineering 2024-07, Vol.178, p.108245, Article 108245
Hauptverfasser: Putzer, M., Zweifel, M., Kneubühler, F., da Silva, G. Rogério, Michael, K., Schröder, N., Schudeleit, T., Bambach, M., Wegener, K.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:•A new calibration method and software is proposed for the self-optimizing calibration of a laser system, for both mechanical and optical axes.•The method and software can be used for automated monitoring and recalibration of the optical axis of a laser system to ensure and maintain a high accuracy.•A new concept of a tool fingerprint for a laser system is introduced, which visualizes the most important beam properties across the scan field.•The difference in removal efficiency depending on the beam shape and the size within the scan field was up to 30 %. Short and ultrashort pulse lasers can ablate the vast majority of materials compared to other machining methods, making them the preferred tool in many niche applications. For the use in precision machining, such laser systems need to reach machining accuracy in the range of 1 µm. Usually, a galvanometric scanner is used in such laser systems to scan the laser beam across the scan field to mark or ablate the various materials. To achieve high accuracy, the overall laser system must be calibrated. However, such calibration is time-consuming and must be monitored at regular intervals and repeated if necessary. In this paper, a method and routine are introduced for self-optimizing calibration, monitoring and recalibration of a laser system. After calibration, the maximal deviation error of the laser system is 3.9 µm within the 30 mm x 20 mm scan field, and the optical axes are capable to match the positioning accuracy of the mechanical axes at 1.11 µm. A beam profiling camera is used to measure the beam position needed for calibration within the scan field. At the same, time other properties of the laser beam such as peak intensity, spot size and ellipticity are mapped within the scan field and is then used as a tool fingerprint of the laser beam across the scan field. Based on this, the influence of the position-dependent beam properties on the removal rate and surface quality is determined.
ISSN:0143-8166
1873-0302
DOI:10.1016/j.optlaseng.2024.108245