Position-sensitive neutron detector with 20 × 20 needle counters

A position-sensitive detector for thermal neutrons was fabricated on the basis of the needle counter. The detector consists of 20 × 20 needle counters with a spacing of 5 mm; the sensitive volume of each needle counter is 4 mm in diameter × 10 mm in length and the filling gas is a mixture of 3He and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2021-10, Vol.1014, p.165708, Article 165708
Hauptverfasser: Isozumi, Yasuhito, Tosaki, Mitsuo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A position-sensitive detector for thermal neutrons was fabricated on the basis of the needle counter. The detector consists of 20 × 20 needle counters with a spacing of 5 mm; the sensitive volume of each needle counter is 4 mm in diameter × 10 mm in length and the filling gas is a mixture of 3He and Q gas (4He + 1%isobutane) with a pressure of 0.3–1.0 MPa. In the position-sensitive detector, a resistor mesh was employed for position encoding by the charge division method. It was found that the encoding method brought about a troublesome distortion in the position image spectrum. In this paper, the performance of the position-sensitive detector is described and the present encoding method is discussed proposing another method without the distortion.
ISSN:0168-9002
1872-9576
DOI:10.1016/j.nima.2021.165708