Thermal neutron measurements with an unpowered, miniature, solid-state device
A prototype neutron detector has been created through modification to a commercial non-volatile flash memory device. Studies are being performed to modify this prototype into a purpose-built device with greater performance and functionality. This paper describes a demonstration of this technology us...
Gespeichert in:
Veröffentlicht in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2021-10, Vol.1012, p.165577, Article 165577 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A prototype neutron detector has been created through modification to a commercial non-volatile flash memory device. Studies are being performed to modify this prototype into a purpose-built device with greater performance and functionality. This paper describes a demonstration of this technology using a thermal neutron beam produced by a TRIGA research reactor. With a 4×4 array of 16 prototype devices, the full widths of the beam dimensions at half maximum are measured to be 2.2×2.1 cm2. |
---|---|
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/j.nima.2021.165577 |