PMT glass window sensitivity to gamma-rays: A digital signal processing approach

A digital data acquisition system together with MATLAB software for data analysis were utilised to measure the pulse specification of the scintillation detector assembly. Besides, the sensitivity of a photomultiplier tube (PMT) window to gamma-rays was studied to discriminate the signals produced as...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2020-03, Vol.957, p.163401, Article 163401
Hauptverfasser: Ghahremani Gol, M., Khakzad, M., Jamili, S., Ghal-Eh, N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A digital data acquisition system together with MATLAB software for data analysis were utilised to measure the pulse specification of the scintillation detector assembly. Besides, the sensitivity of a photomultiplier tube (PMT) window to gamma-rays was studied to discriminate the signals produced as a result of gamma-ray interactions inside the glass window from those originating from the scintillator cell. The result showed that an algorithm based on the time features was capable to evaluate the contributions of these two different signals.
ISSN:0168-9002
1872-9576