Thermal camera performance and image analysis repeatability in equine thermography
•Different imagers were employed to capture standard thermographic views of horses.•Two imagers performed outside of their specifications during quality assurance.•Camera dependent factors contributed to 83% of variance in temperature measurement.•Reliable equine thermography is dependent on appropr...
Gespeichert in:
Veröffentlicht in: | Infrared physics & technology 2020-11, Vol.110, p.103447, Article 103447 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Different imagers were employed to capture standard thermographic views of horses.•Two imagers performed outside of their specifications during quality assurance.•Camera dependent factors contributed to 83% of variance in temperature measurement.•Reliable equine thermography is dependent on appropriate choice of thermal imager.
Infrared thermography is a non-invasive and non-ionizing imaging technique for recording body surface temperature. Considerable research efforts have been made to ensure that thermography in human medicine is performed to consistent standards so that measurements are traceable, reliable and valid, but interest in understanding sources of measurement uncertainty has developed less rapidly in veterinary thermography. The objective of our study was to understand the influence of variability in thermal imager performance, and the contribution to measurement variance of operator-dependent factors, in a practical equine thermography setting. The study employed five different models of thermal imager, which were all quality assured against a blackbody source prior to equine measurements. Three standard thermographic views were then obtained from nine clinically healthy horses using all five thermal imagers, with each imager capturing all views twice. On every thermographic image, regions of interest (ROIs) were then determined by two analyzers. The thermograms from each imager were assessed for their clinical utility. Agreement between analyzers of the same images, and between pairs of repeated images for the same analyzer, was assessed by Bland Altman plots. The contribution to the total variance in temperature measurements from analyzers and from camera-dependent factors was evaluated by ANOVA gauge R&R (reliability and reproducibility) analysis. Two of the thermal imagers produced readings during quality assurance which were outside of their stated specifications for accuracy. There were notable differences in the ability of the imagers to produce clinically useful thermograms, and it was difficult to compare the thermograms from different manufacturers due to the varied color palettes employed. Assessing agreement between analyzers, the bias was −0.05 °C and limits of agreement were 0.11 °C (upper) and −0.21 °C (lower). Agreement between repeated images was less good than between analyzers: the bias was 0.17 °C and limits of agreement were 1.55 °C (upper) and −1.21 °C (lower) for the first analyzer. Across all horses, the contribution |
---|---|
ISSN: | 1350-4495 1879-0275 |
DOI: | 10.1016/j.infrared.2020.103447 |