Evaluation of a nanoflow interface based on the triple-tube coaxial sheath-flow sprayer for capillary electrophoresis-mass spectrometry coupling in metabolomics

•The triple-tube coaxial sheath flow sprayer can be operated in-axis for CE-MS.•The in-axis configuration allows for nanoflow in absence of nebulizing gas.•Sensitivity improvement is function of µeff.•Limits of quantification were improved by a factor of up to six.•Quantification of amino acids was...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Chromatography A 2021-03, Vol.1641, p.461982, Article 461982
Hauptverfasser: Ferré, Sabrina, Drouin, Nicolas, González-Ruiz, Víctor, Rudaz, Serge
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:•The triple-tube coaxial sheath flow sprayer can be operated in-axis for CE-MS.•The in-axis configuration allows for nanoflow in absence of nebulizing gas.•Sensitivity improvement is function of µeff.•Limits of quantification were improved by a factor of up to six.•Quantification of amino acids was achieved in certified plasma. The performance of an original CE-MS interface that allows the in-axis positioning of the electrospray with respect to the MS inlet was evaluated. The variations in the geometrical alignment of this configuration in the absence of a nebulizing gas afforded a significant reduction in the sheath-liquid flow rate from 3 µL/min to as low as 300 nL/min. The sheath liquid and BGE were respectively composed of H2O—iPrOHCH3COOH 50:50:1 (v/v/v) and 10% acetic acid (pH 2.2). A significant gain in sensitivity was obtained, and it was correlated to the effective mobility of the analytes. Compounds with low mobility values showed a greater sensitivity gain. Special attention was paid to the detection of proteinogenic amino acids. Linear response functions were obtained from 15 ng/mL to 500 ng/mL. The limits of quantification, as low as 34.3 ng/mL, were improved by a factor of up to six compared to the conventional configuration. The in-axis setup was ultimately applied to the absolute quantification of four important amino acids, alanine, tyrosine, methionine and valine, in standard reference material (NIST plasma). The accuracies ranged from 78 to 113%, thus demonstrating the potential of this configuration for metabolomics.
ISSN:0021-9673
1873-3778
DOI:10.1016/j.chroma.2021.461982