Unusual effect of flow rate on retention in analytical supercritical fluid chromatography exemplified by polyethylene glycol separation
•Separation of polyethylene glycol oligomers by SFC was studied.•An unusual effect of flow rate on retention was observed.•Retention times at flow rate gradient are lower than at highest constant flow rate.•Effect is observed only starting from a certain molecular weight of PEG oligomers.•The starti...
Gespeichert in:
Veröffentlicht in: | Journal of Chromatography A 2020-01, Vol.1610, p.460513, Article 460513 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Separation of polyethylene glycol oligomers by SFC was studied.•An unusual effect of flow rate on retention was observed.•Retention times at flow rate gradient are lower than at highest constant flow rate.•Effect is observed only starting from a certain molecular weight of PEG oligomers.•The starting molecular weight value depends on chromatography conditions.
We report a case of a peculiar effect of flow rate on retention in a separation of polyethylene glycol oligomers via supercritical fluid chromatography. During method development, we tested flow rate gradients and notices that for some PEG oligomers retention times at flow rate gradient were lower than at constant flow with the largest flow rate value used in a gradient. For instance, at BEH stationary phase and CO2−MeOH gradient from 10 to 35% at 20 min a PEG oligomer having mass of 1225 Da has a retention time 14 min at 1 mL/min flow rate, 10.3 at 2 mL/min and 9.5 min at 1–2 mL/min flow rate gradient. The effect is not unified for all PEG oligomers, it occurs only starting from a particular PEG molecular weight which depends on the stationary phase type and/or mobile phase conditions. We believe that such an unusual flow rate effects can happen in SFC on various occasions, not exclusively for flow rate gradients, and thus should be taken into account during method development or method transfer. |
---|---|
ISSN: | 0021-9673 1873-3778 |
DOI: | 10.1016/j.chroma.2019.460513 |