A discussion on the affecting factors of the fitting procedures’ reliability in isothermal titration calorimetry analysis
The standard isothermal titration calorimetry (ITC) curve, characterized as a typical sigmoid is strictly confined by the so-called c value, which is a ratio of titrand concentration to KD. The proper c value with a range from 5 to 500 is commonly recommended as a standard protocol in routine detect...
Gespeichert in:
Veröffentlicht in: | Archives of biochemistry and biophysics 2021-11, Vol.713, p.109045-109045, Article 109045 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The standard isothermal titration calorimetry (ITC) curve, characterized as a typical sigmoid is strictly confined by the so-called c value, which is a ratio of titrand concentration to KD. The proper c value with a range from 5 to 500 is commonly recommended as a standard protocol in routine detection process for acquiring the reliable fitting results in 1:1 binding mode. However, if the c value is less than “1” due to the weak binding or low concentration of analyte, fitting precision gets unstable and susceptible to the data noise. Herein, we first got a deep discussion into the reliability of the fitting procedure for 1:1 binding mode by data simulation, then quantized the effect of several affecting factors on the precision of parameters estimation through mathematical analysis. Finally, we proposed the value of 2~4 times KD for final ligand concentration is optimal for the ITC titration in low c system (c |
---|---|
ISSN: | 0003-9861 1096-0384 |
DOI: | 10.1016/j.abb.2021.109045 |