X-ray fluorescence analysis of Portland cement and clinker for major and trace elements: accuracy and precision
Accurate, rapid methods for controlling and determining composition are necessary to aid in classification of cements and for correlation of performance characteristics with composition. X-ray fluorescence (XRF) is used to adjust the raw material proportions and to control the process conditions. Th...
Gespeichert in:
Veröffentlicht in: | Journal of the Australian Ceramic Society 2017-10, Vol.53 (2), p.743-749 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Accurate, rapid methods for controlling and determining composition are necessary to aid in classification of cements and for correlation of performance characteristics with composition. X-ray fluorescence (XRF) is used to adjust the raw material proportions and to control the process conditions. The aim of the present study was to establish a reliable analytical program by X-ray fluorescence spectrometry, based on the choice of a well-defined range of standards, which are used to control the quality of the clinker. The results showed a second order polynomial regression model that could properly interpret the experimental data with an
R
2
value >0.97 even after correction for some elements. The results obtained indicate the reliability, reproducibility, and the ease of use of the established program
. |
---|---|
ISSN: | 2510-1560 2510-1579 |
DOI: | 10.1007/s41779-017-0087-x |