Measuring thermal conductivity of powders with differential scanning calorimetry: A simplified method
This paper simplifies a recently proposed method for measuring the thermal conductivity of powders using differential scanning calorimetry (DSC) (Sánchez-Rodríguez et al. in J Therm Anal Calorim 121:469–473, 2015 ). With this method, a crucible is filled with powder and a spherical metal reference i...
Gespeichert in:
Veröffentlicht in: | Journal of thermal analysis and calorimetry 2016-08, Vol.125 (2), p.571-577 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This paper simplifies a recently proposed method for measuring the thermal conductivity of powders using differential scanning calorimetry (DSC) (Sánchez-Rodríguez et al. in J Therm Anal Calorim 121:469–473,
2015
). With this method, a crucible is filled with powder and a spherical metal reference is partially sunk into it. The thermal resistance between the metal and the crucible wall at the metal melting point is obtained from the DSC melting peak slope. In the simplified method outlined in this paper, a cylindrical pan is substituted for the original hemispherical crucible. The equivalence of both methods is demonstrated with alumina powder and commercial cylindrical crucibles of several sizes and aspect ratios. |
---|---|
ISSN: | 1388-6150 1588-2926 |