A liquid scintillation coincidence counter for radiocarbon
Measurements are made of the radiocarbon counting performance of a liquid scintillation coincidence counter used in conjunction with an anticoincidence shield counter. When operated at room temperature, the system has a radiocarbon counting efficiency of 59% at a background of 16 counts/min. The act...
Gespeichert in:
Veröffentlicht in: | Appl. Sci. Research, B B, 1961-03, Vol.9 (2), p.89-92 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Measurements are made of the radiocarbon counting performance of a liquid scintillation coincidence counter used in conjunction with an anticoincidence shield counter. When operated at room temperature, the system has a radiocarbon counting efficiency of 59% at a background of 16 counts/min. The activity of ethanol made from contemporary wood is determined to 13.1 plus or minus 0.6 disintegrations per minute per gram carbon. (auth) |
---|---|
ISSN: | 0365-7140 |
DOI: | 10.1007/BF02930880 |