A sensitive HPLC technique for the quantitation of dauricine

To establish the determination method of dauricine (Dau) concentration in rats' blood and other biological samples, a reverse-phase HPLC method was adopted. Under the given condition, dauricine could be well separated. The retention time (tR) of Dau and its internal standard, daurisoline were 9...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Current medical science 2000, Vol.20 (1), p.39-42
Hauptverfasser: Chen, S, Yang, Y, Zhang, L, Pang, X, Dai, Z, Zeng, F
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To establish the determination method of dauricine (Dau) concentration in rats' blood and other biological samples, a reverse-phase HPLC method was adopted. Under the given condition, dauricine could be well separated. The retention time (tR) of Dau and its internal standard, daurisoline were 9.2 and 6.1 respectively. The detection limit was 10(-2) mg/ml. The absolute recoveries of all kinds of samples were above 70%, and the relative ones were over 85%. A good liner relationship has been obtained over the entire range of 0.030 to 3.000 mg/L in blood samples and 0.050 to 5.000 mg/L in other tissue samples. The intraday and interday coefficients of variation were below 10%. The results showed that the method can be used for detecting Dau in all kinds of biological samples.
ISSN:0257-716X
2096-5230
2523-899X
DOI:10.1007/BF02887672