Methods for measuring the modulation transfer function of gamma camera systems
The MTFs of three different gamma camera systems have been measured using two methods. In one method a step function was used as the object and the calculations were made by means of a pocket calculator. This method has been compared with a conventional line source method, where a computer is needed...
Gespeichert in:
Veröffentlicht in: | European Journal of Nuclear Medicine 1980-01, Vol.5 (1), p.19-22 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The MTFs of three different gamma camera systems have been measured using two methods. In one method a step function was used as the object and the calculations were made by means of a pocket calculator. This method has been compared with a conventional line source method, where a computer is needed in performing the calculations. The methods give the same results with a fairly good accuracy. Both of the methods are well suited to a regular control of the condition and functioning of a gamma camera, provided that the camera is connected to a data-processing system so that the profile curves measured can be converted into the digital form. |
---|---|
ISSN: | 0340-6997 1619-7089 |
DOI: | 10.1007/BF00261201 |