Residual activity of Tc-generators

Owing to contamination by long-lived radio-isotopes, the residual activity of an old 99Mo-99mTc generator may be large enough to prevent its disposal. The long-lived impurities of one generator were determined by energy spectrum and half-life measurements. In this generator 46Sc, 51Cr, 60 Co, 92mNb,...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European Journal of Nuclear Medicine 1981-01, Vol.6 (6), p.269
Hauptverfasser: Väyrynen, T, Näsman, P, Kiviniitty, K
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Owing to contamination by long-lived radio-isotopes, the residual activity of an old 99Mo-99mTc generator may be large enough to prevent its disposal. The long-lived impurities of one generator were determined by energy spectrum and half-life measurements. In this generator 46Sc, 51Cr, 60 Co, 92mNb, 95Nb and 124Sb nuclides were found. Only the half-life of 60Co exceeds 3 months, thus a moderately long storage of the generator reduces the activity, and its disposal is then safe.
ISSN:0340-6997
1619-7089
DOI:10.1007/BF00251351