Microsecond duration pressure pulse generator for in situ dynamically loaded SEM fracture studies
We describe a technique to impulse load small cylindrical specimens for in situ, real‐time SEM studies of dynamically loaded fracture. Loading is done magnetically with a solenoidal coil of 10 mm length and 1 mm radius. The entire loading assembly fits on the SEM stage; there is no need for shock‐is...
Gespeichert in:
Veröffentlicht in: | Scanning 1992, Vol.14 (3), p.169-173 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We describe a technique to impulse load small cylindrical specimens for in situ, real‐time SEM studies of dynamically loaded fracture. Loading is done magnetically with a solenoidal coil of 10 mm length and 1 mm radius. The entire loading assembly fits on the SEM stage; there is no need for shock‐isolated bellows apparatus. The only vacuum feedthrough required is an MHV coaxial bulkhead connection. A ballistic pendulum measurement gave a peak compressive pressure of ∼20 MPa, with a pressure pulse width of 1.33 μs. Pressures > 80 MPa can easily be generated. Using this device we have obtained real‐time SEM movies of dynamically loaded fracture of rosin. |
---|---|