Comparison of Free-Breathing 3D Phase-Resolved Functional Lung (PREFUL) MRI With Dynamic 19 F Ventilation MRI in Patients With Obstructive Lung Disease and Healthy Volunteers
Non-contrast-enhanced H magnetic resonance imaging (MRI) with full lung coverage shows promise for assessment of regional lung ventilation but a comparison with direct ventilation measurement using F MRI is lacking. To compare ventilation parameters calculated using 3D phase-resolved functional lung...
Gespeichert in:
Veröffentlicht in: | Journal of magnetic resonance imaging 2024-10, Vol.60 (4), p.1416-1431 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Non-contrast-enhanced
H magnetic resonance imaging (MRI) with full lung coverage shows promise for assessment of regional lung ventilation but a comparison with direct ventilation measurement using
F MRI is lacking.
To compare ventilation parameters calculated using 3D phase-resolved functional lung (PREFUL) MRI with
F MRI.
Prospective.
Fifteen patients with asthma, 14 patients with chronic obstructive lung disease, and 13 healthy volunteers.
A 3D gradient-echo pulse sequence with golden-angle increment and stack-of-stars encoding at 1.5 T.
All participants underwent 3D PREFUL MRI and
F MRI. For 3D PREFUL, static regional ventilation (RVent) and dynamic flow-volume cross-correlation metric (FVL-CM) were calculated. For both parameters, ventilation defect percentage (VDP) values and ventilation defect (VD) maps (including a combination of both parameters [VDP
]) were determined. For
F MRI, images from eight consecutive breaths under volume-controlled inhalation of perfluoropropane were acquired. Time-to-fill (TTF) and wash-in (WI) parameters were extracted. For all
F parameters, a VD map was generated and the corresponding VDP values were calculated.
For all parameters, the relationship between the two techniques was assessed using a Spearman correlation (r). Differences between VDP values were compared using Bland-Altman analysis. For regional comparison of VD maps, spatial overlap and Sørensen-Dice coefficients were computed.
3D PREFUL VDP values were significantly correlated to VDP measures by
F (r range: 0.59-0.70). For VDP
, no significant bias was observed with VDP of the third and fourth breath (bias range = -6.8:7.7%, P range = 0.25:0.30). For VDP
, no significant bias was found with VDP values of fourth-eighth breaths (bias range = -2.0:12.5%, P range = 0.12:0.75). The overall spatial overlap of all VD maps increased with each breath, ranging from 61% to 81%, stabilizing at the fourth breath.
3D PREFUL MRI parameters showed moderate to strong correlation with
F MRI. Depending on the 3D PREFUL VD map, the best regional agreement was found to
F VD maps of third-fifth breath.
3 TECHNICAL EFFICACY: Stage 2. |
---|---|
ISSN: | 1053-1807 1522-2586 |
DOI: | 10.1002/jmri.29221 |