Large diameter open tubular columns in gas chromatographic analysis
Large diameter open tubular columns provide the packed column chromatographer with a simple route to higher resolution gas chromatography. They can be operated in a high‐flow (lower resolution) mode that permits their direct substitution for a packed column, or they can be operated in a low‐flow (hi...
Gespeichert in:
Veröffentlicht in: | Journal of high resolution chromatography 1986-05, Vol.9 (5), p.272-277 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Large diameter open tubular columns provide the packed column chromatographer with a simple route to higher resolution gas chromatography. They can be operated in a high‐flow (lower resolution) mode that permits their direct substitution for a packed column, or they can be operated in a low‐flow (higher resolution) mode that maximizes separation at the cost of longer analysis times. Inlet design and column installation can influence both the chromatographic results and quantiative reliability. Make‐up gas is not required in the high‐flow mode, and its benefits in the low‐flow mode are restricted to enhanced detector sensitivity, provided the outlet end of the column resides in the detector jet (FID). The columns seem fully compatible with all common modes of detection. |
---|---|
ISSN: | 0935-6304 0344-7138 1521-4168 |
DOI: | 10.1002/jhrc.1240090504 |