High Sensitive Voltammetric Gentamicin Sensor using Poly (3‐thiophenecarboxylic acid‐co‐3‐methylthiophene) Modified Glassy Carbon Electrode

For the determination of gentamicin in biomedical applications, sensitive, fast and low cost voltammetric sensor has been developed. Prepared sensor was successfully used for determination of gentamicine in synthetic urine samples with different voltammetric measurement methods. Firstly, glassy carb...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Electroanalysis (New York, N.Y.) N.Y.), 2022-07, Vol.34 (7), p.1212-1226
Hauptverfasser: Burç, Muammer, Titretir Duran, Serap, Güngör, Öznur, Köytepe, Süleyman
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:For the determination of gentamicin in biomedical applications, sensitive, fast and low cost voltammetric sensor has been developed. Prepared sensor was successfully used for determination of gentamicine in synthetic urine samples with different voltammetric measurement methods. Firstly, glassy carbon electrode was modified using cyclic voltammetry technique by electropolymerization of 3‐thiophenecarboxylic acid and 3‐methylthiophene. Then, optimization of the electrochemical parameters was carried out for modified electrode. With this modified electrode, gentamicin determination was performed in the linear range of 0.05–4.00 mM. The correlation coefficient, limit of detection and limit of quantitation were calculated as 0.9999, 0.039 mM and 0.129 mM, respectively.