Amperometric Detection of Creatinine in Clinical Samples Based on Gold Electrode Arrays Fabricated Using Printed Circuit Board Technology

Herein we report a reliable and stable biosensor for the amperometric detection of creatinine through the ammonium produced in its reaction with creatinine deiminase linked to a Nafion®/polyaniline gold coated surface. The gold electrode array was produced using a low cost printed board circuit tech...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Electroanalysis (New York, N.Y.) N.Y.), 2020-12, Vol.32 (12), p.3054-3059
Hauptverfasser: Ortiz, Mayreli, Botero, Mary Luz, Fragoso, Alex, O'Sullivan, Ciara K.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Herein we report a reliable and stable biosensor for the amperometric detection of creatinine through the ammonium produced in its reaction with creatinine deiminase linked to a Nafion®/polyaniline gold coated surface. The gold electrode array was produced using a low cost printed board circuit technology. A linear range of 1–1000 μM was obtained with standard deviations of less than 5 % (n=3) obtained. Stability studies were carried out and 80 % activity was maintained after one month. The sensors were tested with 142 serum samples from patients with renal dysfunction at different stage of the disease and validated using a commercial colorimetric method.